EasyManua.ls Logo

Keithley 2600a series - Sense Selection; Figure

Keithley 2600a series
696 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2600AS-901-01 Rev. B / September 2008 Return to Section Topics 3-15
Series 2600A System SourceMeter® Instruments Reference Manual Section 3: Basic Operation
Figure 3-3
2-wire resistance sensing
Figure 3-4
4-wire resistance sensing
Sense selection
Front panel sense selection
To select sensing mode:
SourceMeter
HI
LO
I
V
M
Input, Output
V
M
V
R
Test Current (I)
Resistance
Under Test
Lead
Resistances
R
S
R
LEAD
R
LEAD
= Current sourced by SourceMeter
= Voltage measured by SourceMeter
= Voltage across resistor
I
V
M
V
R
Measured resistance = = + (2 X )
Actual resistance = =
V
M
I
R
S
R
LEAD
I
V
R
R
S
SourceMeter
4-wire Sense HI
4-wire Sense LO
I
Input/Output HI
V
M
V
R
Test Current (I)
Resistance
Under Test
Lead
Resistances
R
S
R
LEAD
R
LEAD
= Current sourced by SourceMeter
= Voltage measured by SourceMeter
= Voltage across resistor
I
V
M
V
R
and measured resistance = =
V
M
I
V
M
Sense Current (pA)
Input/Output LO
R
LEAD
R
LEAD
Because sense current is negligible, =
V
M
V
R
I
V
R
=
R
S

Table of Contents