2600AS-901-01 Rev. B / September 2008 Return to Section Topics 4-7
Series 2600A System SourceMeter® Instruments Reference Manual Section 4: Source-Measure Concepts
Figure 4-1
Model 2601A/2602A continuous power operating boundaries
Model 2611A/2635A/2612A/2636A continuous power operating boundaries
The general operating boundaries for Model 2611A/2612A continuous power output are shown in
Figure 4-2 (see General SourceMeter instrument power equation in this section for derating
factors). In this drawing, the 1.5A, 20V and 100mA, 200V magnitudes are nominal values. Also
note that the boundaries are not drawn to scale.
Figure 4-2
Model 2611A/2612A/2635A/2636A continuous power operating boundaries
I-Source operating boundaries
Model 2601A/2602A I-Source operating boundaries
Figure 4-3 shows the operating boundaries for the I-Source. Only the first quadrant of operation is
covered. Operation in the other three quadrants is similar.
-V
+V
-40V
(IV)
Sink
+I
3A
1A
-6V
6V
40V
(III)
Source
(II)
Sink
-1A
-3A
-I
(I)
Source
-2.2A
-600mA
600mA
2.2A
Energy storage DUT
Energy storage DUT
£ 30°C
T
amb
-V
+V
-200V
(IV)
Sink
+I
1.5A
100mA
-20V
20V
200V
(III)
Source
(II)
Sink
-100mA
-1.5A
-I
(I)
Source
Energy storage DUT
Energy storage DUT
£ 30°C
T
amb

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals