19-98 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands Series 2600A System SourceMeter® Instruments Reference Manual
serial
The functions and attributes in this group are used to configure the RS-232 Interface.
serial.baud
Attribute Baud rate for the RS-232 port.
Default 9600
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage baud = serial.baud -- Reads baud rate.
serial.baud = baud -- Writes baud rate.
baud Set to 300, 600, 1200, 2400, 4800, 9600, 19200, 38400,
57600 or 115200.
Remarks • A new baud rate setting takes effect when the command to change it is processed.
• The user should allow ample time for the command to be processed before attempting to
communicate with the instrument again. It is recommended that the baud rate be set from one of
the other command interfaces or from the front panel.
• The baud rate is stored in nonvolatile memory. The reset function has no effect on the baud
rate.
Details See RS-232 interface operation in Section 15.
Also see serial.databits, serial.flowcontrol, serial.parity
Example Sets the baud rate to 1200:
serial.baud = 1200
serial.databits
Attribute Character width (data bits) for the RS-232 port.
Default 8
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage bits = serial.databits -- Reads data width.
serial.databits = bits -- Writes data width.
bits Set to 7 or 8.
Remarks • A new data width setting takes effect when the command to change it is processed.
• The user should allow ample time for the command to be processed before attempting to
communicate with the instrument again. It is recommended that the data width be set from one of
the other command interfaces or from the front panel.
• The data bits value is stored in nonvolatile memory. The reset function has no effect on data
bits.
Details See RS-232 interface operation in Section 15.
Also see serial.baud, serial.flowcontrol, serial.parity
Example Sets data width to 8:
serial.databits = 8

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600a series and is the answer not in the manual?

Keithley 2600a series Specifications

General IconGeneral
BrandKeithley
Model2600a series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals