21-2 Return to Section Topics 2600AS-901-01 Rev. B / September 2008
Section 21: Routine Maintenance Series 2600A System SourceMeter® Instruments Reference Manual
Introduction
The information in this section deals with routine maintenance of the Keithley Instruments Series
2600A System SourceMeter
®
instrument that can be performed by the operator.
Line fuse replacement
WARNING Disconnect the line cord at the rear panel, and remove all test leads
connected to the instrument before replacing the line fuse. Failure
to do so could expose the operator to hazardous voltages that
could result in personal injury or death.
NOTE The power line fuse is accessible from the rear panel, just below the
AC power receptacle (Figure 21-1).
Figure 21-1
Line fuse replacement
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U. S. A .
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
SSLO
LOHI
HIGGG G
CHANNEL B
!
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S . A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
SENSE
LO LO HI
CHANNEL A
SENSE
HI
GUARD
A LO
LAN
NO AUTO-MDIX
SENSE
LO
LOHI
CHANNEL B
SENSE
HI
GUARD
!
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U. S. A .
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
LAN
TSP-Link
R
SENSE
LO LO HI
CHANNEL A
SENSE
HI
GUARD
A LO
!
U
L
C
US
LISTED
SourceMeter
4ZA4
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U. S. A .
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
LAN
TSP-Link
R
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
!
!
Model 2601A/2611A
Line Fuse
Model 2602A/2612A
Line Fuse
Line Fuse
Line Fuse
Model 2635A
Model 2636A