3 Testing 16860 Performance
66 Keysight 16860 Series Portable Logic Analyzer Service Guide
Single Clock Rising Edge - Pod 1 Data - Measuring Maximum Data Rate
1 Select the Run Repetitive icon. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
3 If the clock rate is too high, this error message may occur.
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan rerun
and the test rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute select the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For Single Clock for Rising Edge record the generator frequency and the Data Rate in the
"Maximum State Data Rate" section of the “Performance Test Record. Note: For Single Edge
Clocking, the data rate is the same as the generator frequency.