EasyManua.ls Logo

Keysight Technologies 16861A - Conclude the State Data Rate Tests

Keysight Technologies 16861A
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Testing 16860 Performance
76 Keysight 16860 Series Portable Logic Analyzer Service Guide
Conclude the State Data Rate Tests
Perform the following steps to properly shut down the logic analyzer session after completing the
state mode tests.
End the test.
1 From the Main menu, select File > Exit. In the dialog "Do you want to save the current
configuration?", click No.
2 Disconnect all cables and adapters from the pulse generator.
3 Power down the Analyzer frame.

Table of Contents

Related product manuals