CALibrate:SAVE
Saves the calibration constants in non-volatile memory. Do this at the end of the calibration to avoid
losing changes.
Parameter Typical Return
(none) (none)
Store calibration constants into non-volatile memory: CAL:SAVE
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands. The first parameter specifies the state. The second optional parameter is
the password.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
<password>a numeric value up to 15 digits
Disable calibration: CAL:STAT OFF
Enable calibration with optional password: CAL:STAT ON, 1234
A password is required if <password> has been set to a non-zero value.
CALibrate:VOLTage[:LEVel] <value>, (@<channel>)
Calibrates the output voltage. The value entered selects the range that is being calibrated.
Parameter Typical Return
<value> the maximum voltage rating of the range (none)
<channel> the channel to calibrate
Calibrates the 160 V range: CAL:VOLT 160, (@1)
Keysight MP4300 Series Operating and Service Guide 105
5 SCPI Programming Reference