Parameter Typical Return
<chanlist> one or more channels <49.8,49.9,50.0,50.1,50.2,50.3, ...>
Returns the DAC table settings on channel 2: VOLT:DTAB? (@2)
[SOURce:]VOLTage:DTABle:SASimulator[:IMMediate]? (@<chanlist>)
Returns the DAC table that was calculated based on the SAS parameter settings. The number of
voltage points returned is set at 1024.
Multiple responses are separated by commas. If more than one channel has been queried, all the DAC
values for the first channel are returned, followed immediately by the DAC values of the next channel.
Parameter Typical Return
<chanlist> one or more channels <49.8,49.9,50.0,50.1,50.2,50.3, ...>
Returns the DAC table settings on channel 2: VOLT:DTAB:SAS? (@2)
[SOURce:]VOLTage:DTABle:SASimulator[:IMMediate]:VMP? (@<chanlist>)
Returns the calculated Vmp (maximum power voltage) of the internal DAC table.
Multiple responses are separated by commas. If more than one channel has been queried, all the DAC
values for the first channel are returned, followed immediately by the DAC values of the next channel.
Parameter Typical Return
<chanlist> one or more channels +5.389988E+01
Returns the calculated VMP of the DAC table on channel 2: VOLT:DTAB:SAS:VMP? (@2)
[SOURce:]VOLTage:DTABle:SASimulator[:IMMediate]:VOC? (@<chanlist>)
Returns the calculated Voc (open-circuit voltage) of the internal DAC table.
Multiple responses are separated by commas. If more than one channel has been queried, all the DAC
values for the first channel are returned, followed immediately by the DAC values of the next channel.
Parameter Typical Return
<chanlist> one or more channels +6.000000E+01
Returns the calculated VOC of the DAC table on channel 2: VOLT:DTAB:SAS:VOC? (@2)
[SOURce:]VOLTage:DTABle:TABLe[:IMMediate]? <"name">, (@<chanlist>)
Returns the DAC table that was calculated based on the values from the specified table. The number of
voltage points returned is set at 1024.
Keysight MP4300 Series Operating and Service Guide 143
5 SCPI Programming Reference

Table of Contents