marposs TESTAR E4N - CALIBRATION PROCEDURE; CALIBRATION OF ONE OR MORE MEASUREMENTS, EACH WITH ONE MASTER; CALIBRATION OF ONE OR MORE MEASUREMENTS, EACH WITH TWO MASTERS

marposs TESTAR E4N
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
68
5
PROG
CAL
ENTER
+25.0
PROG
CAL
ENTER
+16.5
A
AIR
PROG
CAL
ENTER
-58.0
PROG
CAL
ENTER
-16.5
A
AIR
PROG
CAL
ENTER
--------
PROG
CAL
ENTER
+16.5
A
AIR
Step 6: Remove the Min. master and insert the Max. master in the measuring position again.
Step 3: Insert the Max. master in the measurement position. Turn the sensitivity adjuster until the value
shown on the alphanumeric display is close to
Step 4: Remove the Max. master and insert the Min. master in the measuring position.
Step 3
Step 5:
Step 7: Turn the sensitivity adjuster until the value shown on the alphanumeric display is close
(Upper limit).
Step 8: Remove the Max. master and insert the Min. master in the measuring position again.
Step 7
Step 5
ADJUSTMENT PROCEDURE