86
10








ERR 202
E 202
ERR 504
ERR 505

ALPHANUMERIC
DISPLAY
7-SEGMENT DI-
SPLAY
ERROR TYPE DESCRIPTION MENU ACTION
PROGRAMMING ERROR MESSAGES AND SOLUTIONS
Programming
The E2PROM is incom-
patible with the column
programming
Press ENTER to reset the
column; re-program the
column.
Hardware
RS232= error, HAND_
SH= incorrect for the
current printer
Press ENTER to reset the
column and check that
the HAND_SH= variable
is correctly programmed
Hardware
Software
Communications error
between CPU and the
Relay-BCD card

and then ON again.
If the problem persists,
call a Marposs Service
Centre.
Hardware
E2PROM checksum
error

and then ON again; re-
program the column.
If the problem persists,
call a Marposs Service
Centre.
Hardware
RAM error

and then ON again.
If the problem persists,
call a Marposs Service
Centre.
Calibration
During calibration of
the two masters, the
automatically calculated
sensitivity adjustment
has exceeded the per-

the difference between
the masters)
Check the masters for
wear. Check that the
masters are correct for the
current type of gauge.
If the problem persists, call
a Marposs Service Centre.
Programming
The measurement value
is outside the LED scale
range or is incompatible
with the LED scale
Modify the scale, the
measurement coefficient
or the nominal value
COMM:
(ADVANC.:)
MEASURE:
MEASURE:
Hardware
Error in the LED scale
configuration
Contact a Marposs
Service Center
Software
System software error
(The current measure-
ment ID is incorrect).

and then ON again.
If the problem persists,
call a Marposs Service
Centre.