System Errors
Current system errors appear at the top of the screen in white letters on a red background. The message
disappears once the error is corrected.
If a system error occurs during standby mode, a warning message appears as the fault occurs but disappears
as the situation is corrected.
If the error occurs during a disinfection cycle, the program ends and uids are pumped out of the basin.
A recovery program runs until a safe state is reached and the endoscope can be removed. The recovery
program can not be stopped by the operator.
System Information
The bar at the bottom of the screen provides system communication information, machine identication,
and the software version.
MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor software has two possible
system connections:
1. MEDIVATORS ADVANTAGE PLUS System PC to the MEDIVATORS ADVANTAGE PLUS
System (SCU)
2. MEDIVATORS ADVANTAGE PLUS System PC to an external server (optional)
The two left-most elds on the bar indicate if the MEDIVATORS ADVANTAGE PLUS System is
connected to the database. The next eld to the right indicates if the PC is correctly connected to the
MEDIVATORS ADVANTAGE PLUS SCU. The round indicator lights are green when a connection has
been established and gray when no connection exists.