110
Server disconnected. Selection aborted. C-015
Causes: The connection to the database has been interrupted.
The LIO program cannot retrieve the overview lists and the selection screen is empty.
The connection to the server is interrupted in stand-by mode.
It is not possible to start a program.
Action: Report the error to the system manager. As long as “Disconnected from server” appears at the
bottom of the screen, the program can not be restarted.
Run could not be congured, check SQL-server connection C-016
Causes: The system is unable to retrieve information from database to set up a run.
Action: Report the error to the MEDIVATORS
®
ADVANTAGE PLUS
®
system manager.
User data could not be retrieved, check SQL-server connection C-017
Causes: The system is unable to retrieve information from database to set up a run.
Action: Report the error to the MEDIVATORS ADVANTAGE PLUS system manager.
The clock has been synchronized with the server computer (%s) C-019
Cause: If the MEDIVATORS ADVANTAGE PLUS System clock and the server clock have a time
difference of more than 30 seconds, the check will be synchronized.
Action: No action is required.
The Expert control has been automatically closed, you have to log on again C-021
Cause: The Expert screen has timed out.
Action: To continue using the Expert functions login again.