111
The previous run has ended incorrectly. Program aborted to a safe situation. C-030
If a system error occurs, then it is not possible to create a correct run report. If the error is repaired, then
the error message disappears and the recovery program automatically runs to rinse the endoscope with
water so that it can be safely removed of the machine. This program can not be stopped and the endoscope
is not disinfected.
Cause: A black screen means the program is not visible. Possible causes are as follows:
• The PC was switched off.
• There is a PC defect.
• There was a power failure.
Action: Try pressing the On/Off switch on the PC. If this does not solve the error, contact the maintenance
department.
Cause: There is a communication problem so the program cannot be run correctly. Possible causes are
as follows:
• There is a cable fault.
• A cable is loose.
• There is a program error.
• The MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor control
unit is defective.
• There is a blown or defective fuse.
Action: Contact the maintenance department and the system manager.
Cause: There is a communication problem with the SCU interface and message W-701 is also on the
screen. Possible causes are as follows:
• There is a fault with the SCU interface.
• A network cable is disconnected.
• A cable is faulty.
• There is a communication problem with the server.
• There is a server error.
Action: Check the cable connection between the network and the machine. Restore the connection if
possible and start the machine program again. If you are unable to repair the connection, report
the error to the system manager
Cause: The LIO program was interrupted. Message W-703 also appears on the screen and the cover is
locked.
Action: Start the recovery program.