112
A failure occurred during the execution of a run. Program aborted to a safe
situation.
C-032
If an SCU error occurs, then it is not possible to carry out a correct disinfection cycle. If the error is
repaired, then this C-032 message appears and the recovery program is automatically selected. The
endoscope is rinsed with water so that it can be safely removed from the MEDIVATORS
®
ADVANTAGE
PLUS
®
Automated Endoscope Reprocessor but it will not be disinfected. The recovery program works
automatically and can not be stopped.
The endoscope must be run through a complete disinfection program before use.
Cause: There is an internal communication problem with the PC. Possible causes are as follows:
There is a cable fault.
A cable is loose.
There is a program error.
The SCU interface in the MEDIVATORS ADVANTAGE PLUS system is defective.
A fuse is blown or defective.
Action: Report the problem to the maintenance department.
Cause: A new program was started while another one was already active.
Action: Press the Start button to start the recovery program.
Program aborted to a safe situation, but incorrect parameter set C-033
Cause: An incorrect parameter set was selected for the machine disinfection or the recovery program.
Action: Select the correct parameter set for the recovery program.
Cause: The LIO program was interrupted or shut down during a run. Message W-703 also appears and
the lid can not be opened.
Action: Start a recovery program.
Hardware denition contains an invalid signal %d C-040
Cause: An incorrect address is entering the controller from the SCU.
Action: Report the error to the maintenance department. Note the unknown signal number: nn.
Selected hardware denition not correct for “%s” type C-042
Cause: The system settings parameters for the left/right unit type have not been selected.
Action: Report the error to the MEDIVATORS ADVANTAGE PLUS System manager.
Automatic start of program is not possible in the current state C-052
Cause: There is warning text in the report eld.
Action: Ensure that all error messages are resolved. Select a machine disinfection auto start. If necessary,
consult the MEDIVATORS ADVANTAGE PLUS System manager.
Cause: An auto start for endoscope disinfection is already set up.
Action: Cancel the auto start for the endoscope disinfection.