Error Number Description Additional Details
F041 Air channel To control ow, the relevant channel sensor must read high or
low for a minimal period of time. A “high” error occurs if a pump
or valve is broken, or if the sensor is defective or blinks. A “low”
error may be caused by a partially-blocked channel or terminal
block.
F042 Biopsy channel
F043 Jet channel
F044 Suction channel
F045 Water channel
F046 Extra channel
F047 Elevator channel
F051 Basin empty The basin empty sensor does not detect the rinse basin
draining. If there is water in the basin, the sensor is defective
or the drainage hose is blocked. If when adding disinfectant,
the basin empty sensor does not go on, the sensor is out of
alignment, is defective, or the endoscope housing is blocking
the ow of disinfectant.
F052 Minimum level The basin full sensor is not activated when the basin lls. The
sensor is contaminated or the water supply is too slow. During
rinsing the basin full level sensor reads incorrectly due to a
valve leak.