128
Error Number Description Additional Details
F053 Dosing reservoir A not
empty
Part A dosing reservoir not emptied at end of delivery
F054 Dosing reservoir B not
empty
Part B dosing reservoir not emptied at end of delivery
F055 Dosing reservoir A not full Part A dosing reservoir is not full at start of delivery
F056 Dosing reservoir B not full Part B dosing reservoir is not full at start of delivery
F057 Dosing pump A Part A ll pump is on during delivery
F058 Dosing pump B Part B ll pump is on during delivery
F062 Detergent The relevant sensor is activated when it is not applicable in
the step. When adding disinfectant or detergent, a message
appears if the minimal dosage time is not achieved or the
maximum dosage time was exceeded.
F072 Pump detergent The pump runs at the incorrect time due to a software or
electronics error.
F073 Pump channels
F074 Pump alcohol Alcohol pump high during step
F075 Minimum spray arm speed Spray arm spinning less than 15 rpm
F076 Maximum spray arm speed Spray arm spinning more than 250 rpm
F081 Drainage valve The valve is open at the wrong step due to a software or
electronics error.
F082 Water inlet
F091 Volume change The volume meter counts pulses at the wrong step, possibly
due to a leaking supply valve.
F092 Volume minimum When lling the rinse basin, the measured volume is less than
the minimum or more than the maximum setting.
F094 V1 > V2 During lling the V2 meter veries the amount measured by V1.
The deviation between the two meters is greater than 15%.
F095 V1 < V2
F101 Max temperature exceeded The water is warmer than the upper limit (40°C) or colder than
the lower limit (30°C). Water temperature adjustment is done
externally.
F102 Temperature below
minimum
F103 Basin temperature Temperature sensors differ by more than 3ºC