129
Cause: The Safety Control (SCU) program detects an error. A sensor or valve may have had an
incorrect status for a short period of time so the program was rejected as unsafe. If the fault
occurred at the end of a step, the message is in the run report of the next step.
Action: A correct description is important for a good analysis.
• Note the codes and the text.
• Report the fault and description to the maintenance department.
• Wait until the program has stopped.
• Press the Open/Close
button.
• Start a restore program.
• Restart the endoscope disinfection.
Cause: A PC error disrupts the program. The ASC signals the unsafe situation and sends an error
message.
Action: Note the code and description. Evaluate if the PC is working. Report the fault and description
to the maintenance department.
SCU in “OK” state S-301
SCU : Wrong conguration S-302
The PC communicates with the SCU operation interface in the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor. This interface transmits a signal every 20 msec to check the status of
the signal in and out of the PC. If this communication stops or is interrupted the error appears.
Cause: The safety control program found a fault in one of the electronic components or the controller
is not running.
Action: Report the fault to the maintenance department.
SCU : no connection S-304
Cause: A communication error occurred between the PC and the electronics drive controller.
1. No serial connection with the PC
2. The power or plug is disconnected
3. The drive controller is defective.
Action: Make certain all cables are connected to the PC and the 230V plug. Report the fault to the
maintenance department.
Manual control active S-350
Cause: The manual control in the Expert screen is active and has not been released.
Action: Click the “release all” button on the Expert screen.