132
Lid will not close W-713
Action: Press the Start
button to close the cover or close the cover carefully by hand and test to
see if it moves.
Cause: The cover is blocked.
Action: See if the blockage can be seen and attempt to remove it if possible. If no blockage can be seen,
report the problem to the maintenance department.
Cause: The cover sensor is defective or out of alignment.
Action: Consult the maintenance department.
Cause: The Open/Close
button was pressed down too long.
Action: Press the button again but for a shorter time.
Cause: There is a defective air valve or the compressed air pressure is too low to open the cover.
Action: Report the situation to the maintenance department.
Basin not empty W-715
Cause: The program was aborted and there is still uid in the basin.
Action: Open the cover to shut down the program and automatically drain the basin.
Cause: There is liquid in the basin because the drain is blocked or the valve is defective.
Action: Check if the drain is used and report the problem to the maintenance department.
Cause: There is no liquid in the rinse basin. There is a basin sensor defect.
Action: Report the problem to the maintenance department.
Fluid leak detected W-716
Cause: Liquid was spilled while changing a reservoir.
Action: Remove the liquid and dry the sensor (the green plug) carefully. Resetting the sensor can take
a few minutes.
Cause: While in stand-by, liquid is detected at the bottom of the machine. Possible causes are leaking
or a sensor set at the wrong level
Action: Report the situation to the maintenance department.
Water temperature sensor defective W-721
Cause: There is a temperature sensor defect if the reading is >100°C.
Action: Report the problem to the maintenance department.
Sensor S03 Detergent high W-743
Cause: A sensor is out of alignment or the detergent dosage system is clogged.
Action: Report the warning to the maintenance department.