Sensor S17 Minimum level high W-744
Cause: There is water in the rinse basin or a sensor defect.
Action: Close the cover to allow the basin to run empty. Report the warning to the maintenance
department.
Sensor S24 : Alcohol high W-749
Cause: A sensor is out of alignment or the alcohol dosage system is clogged.
Action: Report the warning to the maintenance department.
Sensor S11 Leak test #1 high W-751
Cause: Disinfection cycle fails or the user pushed the cancel button to end the cycle
Action: Open the basin lid and disconnect the channel block from the channel manifold. The pressure
should now be released and the warning message should disappear. Reinstall the channel block,
close the basin lid, and a new cycle can now be initiated.
Sensor S25 high W-753
Cause: The part A dosing reservoir ll sensor is out of alignment or faulty
Action: Report the warning to the maintenance department.
Sensor S26 high W-754
Cause: The part A dosing reservoir ll sensor is out of alignment or faulty
Action: Report the warning to the maintenance department.
Sensor S33 high W-757
This message appears if not disinfection program is active. The warning must be cleared before a program
is restarted.
Cause: The air cylinder sensor is out of alignment or faulty.
Action: Report the error to Maintenance.