134
Sensor high
W-764, W-765,
W-766, W-767,
W-768, W-769,
W-770
This message appears when the machine is in stand-by and one of the channel sensors is high. This report
occurs for the following channels:
• Air • Suction
• Biopsy • Jet
• Extra • Elevator wire
• Water
Cause: The connection block is obstructed and the endoscope is not connected.
Action: Disengage the block. If the message disappears, report the obstructed block to the maintenance
department. If the message remains see the next possible cause.
Cause: There is a channel sensor defect.
Action: Report the defective sensor to the maintenance department.
Manual control active W-800
Cause: One or more functions are manually set. This means a disinfection program can not be started.
Action: Contact the maintenance department with a request to cancel the setting.
Purge in progress W-801
Cause: This message appears when the machine is in stand-by and the machine leak test system is
blown through. This is an automatic routine that occurs every hour and lasts one minute.
Action: Wait until the message disappears and start the program.
Venting alcohol line, close lid W-802
Cause: Alcohol lines are pressurized while unit is idle.
Action: Wait until venting process completes, then start cycle.
Discharging air cylinder, close lid W-803
Cause: Air cylinder is pressurized while unit is idle.
Action: Wait until venting process completes, then start cycle.
Left side controls the water line disinfect W-805
Cause: Right side cannot control the water line disinfect cycle.
Action: Use left side control panel to control water line disinfect cycle.