135
Open lid to complete cycle W-810
Cause: The program is nished. This report indicates how the program needs to be shut down.
Action: Press the Open/Close button and follow the instructions on the screen. After selecting
the correct operator ID, the cover opens automatically and the program shuts down.
Disinfectant in the basin (lid locked) W-820
Cause: The water line disinfect cycle failed while disinfectant was in the basin.
Action: Restart the water line disinfect cycle.
Disinfectant in supply line (lid locked) W-821
Cause: The water line disinfect cycle failed while disinfectant was in the water supply lines.
Action: Restart the water line disinfect cycle.