Disinfectant
Rapicide PA Part A and Part B
The two disinfectant containers serve both the left and right modules. One container holds Part A while
the other container holds Part B of the disinfectant solutions. The MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor automatically mixes the two parts with water at the appropriate time
during the disinfection cycle.
Note: Upon replacing RAPICIDE
®
PA Part A and B containers, the Replace Fluids utility must be run.
A reprocessing cycle cannot be initiated if an out of disinfectant has been detected. The Replace Fluids
utility must be run to clear the error condition.
Figure 26 Parts A and B
Part A uses a blue cap, and is positioned to the left as you look into the front of the MEDIVATORS
ADVANTAGE PLUS System. Part B uses a white cap, and is positioned to the right as you look into the
front of the MEDIVATORS ADVANTAGE PLUS System. Part A and B must always be positioned in the
correct location in the machine, and must always be attached to the appropriately colored cap. To facilitate
the correct connection, the blue cap is larger than the white cap, and only ts the Part A container. The
smaller white cap only ts the Part B container.
Contact Time
Prior to replacing Part A or B the Replace Fluids utility must be run. Information from the containers is
required. The disinfectant contact time and disinfectant temperature is pre- set during manufacturing. The
contact time and temperature eld is not congurable.