93
Water lters should be replaced at a minimum every three months for the 1.0 micron and 0.4 micron lters
and every six months for the 0.1 micron lter. If the water pressure drops below 40 PSI or the difference
between the gauges is greater than 5 PSI the corresponding lter needs to be replaced.
The following conditions must be met before changing a lter:
No endoscope reprocessing cycles are running.
No water line disinfection procedure is running.
The inlet water supply is turned off.
You will need the following materials:
A container to catch water.
A lter bowl wrench.
The necessary lter cartridges (see Table 1) (See Appendix B for replacement part numbers).
Table 1 Filter Cartridges
Key
Filtration
Porosity
Filter Type
Minimum Replacement
Schedule
1 0.1 micron Hollow ber 6 mos.
2 0.4 micron Pleated 3 mos.
3 1 micron Pleated 3 mos.
Figure 5 Filter Cartridges