94
Figure 6 Filter Cartridges in place
1. Relieve the pressure in the ltration system by opening the bleeder valve on the lter system (rotate
down).
2. Check the inlet pressure gauge to ensure that the pressure is at zero.
3. Place an empty container under the 1 micron lter housing and disconnect the Quick Connect tting
at the bottom of the housing. Allowing all of the water to empty out.
4. Using a lter bowl wrench, remove the 1 micron lter housing and discard the used lter cartridge.
Figure 7 Loosen Filter Housing