96
Figure 10 Open Bleed Valve
8. To remove the 0.1 micron lter rotate the used lter cartridge counterclockwise, pull downward,
and discard.
9. To install a new 0.1 micron lter insert a new lter cartridge into the housing header and rotate
clockwise until the tabs lock into place gure 11.
Figure 11 Filter Mounting Tabs
10. Return the bleed valve to the closed position (rotate up).
11. Slowly open the main water supply valve and check for any leaks.