12. Run the Water Sample Utility
• Push Menu button
• Log in
• Select: Program
• Select: Utility Program
• Choose: Take Water Sample
13. Place a container under the bleed valve outlet tube and slowly open the bleed valve (13 g. 4). Let the
water run until it is a steady stream.
14. Close the bleed valve.
15. Push cancel to stop water utility or allow to complete. No actual sample needs to be taken.
16. Perform a water line and water ltration disinfection procedure. Refer to Chapter 2.