EasyManua.ls Logo

MICRO-EPSILON ILD 1420-10 - A 3.3.11.4 BASICSETTINGS, Load; Save Device Settings; A 3.3.11.5 SETDEFAULT, Default Settings; A 3.3.12 ANALOGSCALE, Scaling the Analog Output

Default Icon
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 105
Appendix| ASCII Communication with Sensor
optoNCDT 1420
PRESETMODE: Returns the set signal quality
PRESETMODE <Mode>: Setting the signal quality. Setting the signal quality is possible only, if a preset
was loaded.
<mode> = STATIC|BALANCED|DYNAMIC|NOAVERAGING|NONE
A 3.3.11.4 BASICSETTINGS, Load / Save Device Settings
BASICSETTINGS READ | STORE
- READ: Loads the stored device settings from the senosr.
- STORE: Saves the current device settings in the sensor.
A 3.3.11.5 SETDEFAULT, Default Settings
SETDEFAULT ALL | MEASSETTINGS | BASICSETTINGS
Sets sensor back to default settings.
- ALL: Measurement and device settings are being deleted. The standard preset for the measurement set-
ting resp. the the default parameter for the device settings are being loaded.
- MEASSETTINGS: Measurement settings are being deleted and the standard preset is being loaded.
- BASICSETTINGS: Device settings are being deleted and the default parameters are being loaded.
A 3.3.12 ANALOGSCALE, Scaling the Analog Output
ANALOGSCALE STANDARD| (TWOPOINT <Minimum value> <Maximum value>)
Setting the two point scaling of the analog outpt.
- STANDARD: using the measuring range of the sensor
- TWOPOINT: two point scaling within the analog range (4 - 20 mA)
Minimum value: measurement value in mm which is matched to the lower analog value (4 mA)
Maximum value: measurement value in mm which is matched to the upper analog value (20 mA)
i
The minimum value (in mm) can be higher than the maximum value (in mm), see Chap. 7.5.3.

Table of Contents