EasyManua.ls Logo

MICRO-EPSILON ILD 1420-10 - Triggering; General

Default Icon
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 52
Set Sensor Parameter
optoNCDT 1420
7.4.4 Triggering
7.4.4.1 General
The optoNCDT 1420 measurement input and output is controllable through an external trigger signal or a
command. Triggering affects the analog and digital output. The measurement value at the time of triggering is
output delayed, see Chap. 6.4.
- Triggering does not influence the measuring rate resp. the timing so that between the trigger event (level
change) and the start of output always lie 3 cycles + 1 cycle (Jitter).
- Micro-Epsilon recommends the abdication of data reduction, for example, by sub-sampling when the
triggering is used.
- The multifunctional input is used as external trigger input, see Chap. 5.4.6.
- Factory setting: no triggering, the sensor starts data transmission right after start-up.
- Pulse duration of the “Trigger in“ signal must be at least 50 µs.
Triggering of measured value acquisition and output have the same time behavior.
Input trigger
/
Output
trigger
Level A continuous measurement task is following as long as
the chosen level remains the same. Choice of level, see
Chap. 7.3. Pulse duration must be at least a cycle time.
The following pause must be at least a cycle time..
Edge infinite Edge selection, see Chap. 7.3. “0“ end trigger,
“1 ... 16382“ values per trigger, “16383“ endless trigger
manual Number Value
Software infinite A software triggering is started by clicking the button Re-
lease trigger. “0“ end trigger
“1 ... 16382“ values per rigger, “16383“ endless trigger
manual Number Value
Inactive No triggering
Grey shaded
fields require a
selection.
Value
Dark-bordered
fields require you
to specify a value.

Table of Contents