Page 67
Set Sensor Parameter
optoNCDT 1420
7.5.2.2 Characteristics Digital Output
Measurements are coded with 18 bit, if the based on zero setting or mastering. The master value itself can accept the double measu-
ring range. The examples below show the digital output behavior of an ILD1420-50 with 50 mm measuring range.
Target at 16 % measuring range Target at 60 % measuring range Target at 60 % measuring range
SMR EMR
0 % 16 % 100 % MR
Target
10920 8.00 mm
SMR EMR
0 % 60 % 100 % MR
Target
3918330.00 mm
SMR EMR
0 % 60 % 100 % MR
Target
3918330.00 mm
Zero setting (master value = 0 mm) Zero setting (master value = 0 mm) Set master value 10 mm
SMR EMR
0 % 16 % 100 % MR
Target
32760 0.00 mm
22483
-8.00 mm
86717
42.00 mm
EMR
0 % 60 % 100 % MR
Target
327600.00 mm
10 %
643
-30.0 mm
58454
20.00 mm
EMRSMR
0 % 60 % 100 % MR
Target
4560710.00 mm
7066
-20.0 mm
71301
30.00 mm
Digital minimum reached at 10 % MR