Appendix| ASCII Communication with Sensor
optoNCDT 1420
A 3.7 Error Messages
If an error occurs with a command, the error message is listed.
Error message Description
E100 Internal error Internal error code
E104 Timeout Timeout while mastering.
E200 I/O operation failed Cannot write data to the output channel.
E202 Access denied Access denied: Login as expert is necessary.
E204 Received unsupported character An unsupported character was received
E210 Unknown command Unknown command rights to small to read).
E214 Entered command is too long to be
processed
The entered command with the parameters is too long
(greater than 255 bytes).
E220 Timeout, command aborted Timeout during mastering.
E232 Wrong parameter count Too high or too small number of parameters.
E234 Wrong or unknown parameter type A transmitted parameter has a wrong type or a wrong num-
ber of parameters were transmitted.
E236 Value is out of range or the format is
invalid
The parameter value is out of range of the value range.
E262 Active signal transmission, please stop
before
A measurement value transmission is active. Stop the data
transmission in order to execute the command.
E320 Wrong info-data of the update For update only: the header of update data contains an
error.
E321 Update file is too large For update only: update data is too large.
E322 Error during data transmission of the
update
For update only: error during update data transmission.
E323 Timeout during the update For update only: Timeout during the transmission of update
data.
E331 Validation of import file failed The import file is not valid.