To fill the reservoir, do these steps:
1. Remove the reservoir from the package, and fully extend the plunger.
plunger
reservoir transfer
guard
1
2. Swab the vial with alcohol (not shown).
3. Press the transfer guard onto the vial without pushing down on the plunger.
vial
3
4. Push down on the plunger to pressurize the vial. Hold down the plunger rod.
5. While still holding down the plunger rod, flip the vial over so the vial is on
top. Slowly pull down on the plunger to fill the reservoir.
6. Gently tap the side of the reservoir to make any air bubbles rise to the top of
the reservoir.
6
5
4
reservoir and infusion set
Reservoir and infusion set 97