Title and text Explanation Next steps
Active Insulin cleared
Any Active Insulin amount
has been cleared.
An alarm has
cleared your active
insulin amount.
Your active insulin
amount is now at
0 units.
• Select OK to clear the alarm.
• The active insulin tracked prior to
pump restart is not included in
new Bolus Wizard calculations.
Consult your healthcare
professional for how long you
need to wait after active insulin is
cleared before you can rely on
the active insulin calculation of
your Bolus Wizard
• You can check Daily History for
the time and amount of your last
bolus. For more information, see
Daily History, on page 121.
Auto Suspend
Insulin delivery
suspended. No buttons
pressed within time set in
Auto Suspend.
You have not
pressed any buttons
during the time
specified in the
Auto Suspend
settings.
• To clear the alarm and resume
basal insulin delivery, select
Resume Basal.
• Check your BG and treat as
necessary.
Battery failed
Insert a new AA battery.
The pump battery
does not have
enough power.
• Select OK to clear the alarm.
• Remove the old battery and
insert a new AA battery.
For details, see About batteries, on
page 21.
Battery not compatible.
See User Guide.
The battery that you
inserted into the
pump is not
compatible.
• To clear the alarm, remove the
incompatible battery.
• Insert a new AA battery. For
compatible battery types, see
About batteries, on page 21.
■
alarms, alerts, and messages
Alarms, alerts, and messages 203
MP6025957-024 / A
RELEASED