Title and text Explanation Next steps
Bolus not delivered.
Bolus entry timed out
before delivery. If bolus
intended, enter values
again.
Bolus values
entered, but bolus
was not delivered
within 30 seconds.
Select OK to clear the alert.
If bolus delivery was intended,
check your BG, re-enter bolus
values and deliver bolus.
Bolus stopped
Cannot resume bolus or
cannula fill. XX.XXX of
YY.YYY U delivered.
ZZ.ZZZ U not delivered. If
needed, enter values
again.
The battery power
was exhausted
while a bolus or Fill
Cannula was in
progress.
Note the amount of insulin not
delivered.
Replace the AA battery.
Select OK to clear the alarm.
Deliver the remaining bolus
amount if needed.
Cannot connect device
This device is
incompatible with your
pump. See User Guide.
You may be trying
to connect a
device that is not
compatible with
your pump.
You are trying to
connect a
transmitter to your
pump but another
transmitter is
already wirelessly
connected to your
pump.
Select OK to clear the alert.
Check the list of devices that are
compatible with your pump in
Optional devices, on page 14.
If you are replacing your
transmitter, make sure that you
first delete the old transmitter
from your pump before you try
to connect your new transmitter.
Only one transmitter can be
connected to your pump. For
details, see Deleting the
transmitter from your pump, on
page 181.
Check settings
Startup Wizard settings
complete. Check and set
up your other settings.
Some settings have
been cleared or
reverted to factory
default values.
Select OK to clear the alert.
Review any settings that you
have not already set in Startup
Wizard and re-enter the values, if
necessary.
204 Chapter 12
MP6025957-024 / A
RELEASED