Title and text Explanation Next steps
Critical pump error
Delivery stopped. Pump
not working properly.
Stop using pump.
Remove infusion set from
body. Consider other
insulin treatment. See
User Guide.
Your pump has
encountered a
critical error.
The pump is not able to deliver
insulin. Remove your infusion set
and stop using your pump.
• Consider another form of insulin
delivery.
• Check your BG, and treat as
necessary.
• Write down the error code that
appears on the alarm screen.
• Call your local representative for
assistance with your pump.
Delivery limit exceeded
Delivery stopped. Check
BG. See User Guide for
more information.
You have attempted
to deliver more
insulin than
expected based on
your Max Bolus and
Max Basal settings.
• Check your BG.
• Select Resume Basal.
• Check Bolus History and
reevaluate your need for insulin.
• Continue to monitor your BG.
Fill Cannula?
Select Fill to fill cannula
or select Done if not
needed.
You had the Fill
Cannula screen
displayed for
15 minutes.
• To proceed and fill the cannula,
select Fill.
• If you do not need to fill the
cannula, select Done to skip this
process.
Insert battery
Delivery stopped. Insert a
new battery now.
The battery was
removed from the
pump.
• Insert a new AA battery.
• The alarm clears when you insert
a new battery.
• The pump powers off after
10 minutes unless you insert a
new battery.
■
alarms, alerts, and messages
Alarms, alerts, and messages 205