Title and text Explanation Next steps
Insulin flow blocked
Check BG. Consider
injection and testing
ketones. Estimated 0 U
insulin in reservoir.
Change reservoir and
infusion set.
Your pump has
detected that the
insulin flow was
blocked and there is
no insulin in the
reservoir.
Check your blood glucose.
Consider checking ketones and
take an injection if needed.
Remove your infusion set and
reservoir.
Select Rewind to start the new
reservoir process using a new
infusion set and reservoir.
If a bolus delivery was in progress
when the alarm occurred:
Check the Daily History screen for
the amount of bolus already
delivered before the pump
alarmed.
Consider delivering remaining
bolus, if the bolus insulin was not
included in an insulin injection.
alarms, alerts, and messages
Alarms, alerts, and messages 207
MP6025957-024 / A