Title and text Explanation Next steps
Insulin flow blocked
Fill Cannula stopped.
Remove infusion set from
body. Change reservoir
and infusion set.
Your pump has
detected the insulin
flow was blocked
while filling the
cannula.
Check your blood glucose.
Consider checking ketones and
take an injection if needed.
Remove your infusion set and
reservoir.
Select Rewind to start the new
reservoir process using a new
infusion set and reservoir.
If a fill cannula was in progress
when the alarm occurred:
Check the Daily History screen for
the amount of bolus already
delivered before the pump
alarmed.
Consider delivering remaining
bolus, if the bolus insulin was not
included in an insulin injection.
208 Chapter 12
MP6025957-024 / A
RELEASED