36
3 Advanced Operations and Useful Functions
No. 99MAM029A
y In MP mode
LOG OUT1 LOG OUT2 LOG OUT3
Measurement
date and time
Measurement value
Data number
Data input axes
Measurement value
Data input axes
Measurement time
Mea-
surement
date
1
Set "LOG RESUME" and "OUT LOG" of the parameters.
1
While pressing
DATA
press
POWER
,
then just release
POWER
, when the print-
ing starts, then release
DATA
.
» The power is turned on and enters the
parameter setup mode.
2
Press
DATA
repeatedly until "17 LOG
RESUME" is printed.
» "17 LOG RESUME : 1" is printed.
3
Press
STAT.
OUT LOG
repeatedly until the de-
sired number to set to LOG RESUME is
printed.
4
Press
DATA
.
5
Press
STAT.
OUT LOG
repeatedly until the de-
sired number to set to "18 OUT LOG" is
printed.
6
Press
DATA
.
» The parameter settings list is printed.
» The parameter settings are completed.