45
3 Advanced Operations and Useful Functions
No. 99MAM029A
3.6.2 Deletion of the Measurement Data
Although
CLEAR
is pressed to delete the measurement data stored in the product inter-
nally, the deletion target may vary according to the WORK MODE as shown below.
MODE0, MODE1, MODE2
MODE3
During subgroup
measurement
After completion of
subgroup measure-
ment
Only the measurement data is deleted.
Press it before tolerance setting (limit
data setting) without fail.
Re-input from the
No.1 data.
Delete all the measure-
ment data.
When deleting the log data, hold down
CLEAR
while pressing
TOL.
REC/STOP
.
For further details on the logging operation of measurement data, refer to
"3.4.2
Start/Stop of Logging, Collective Print/Deletion of Log Data" on page 37.
Tips
• The parameter setting is not changed even if
CLEAR
is pressed. When initializing the parame-
ters, select "PARAMETER CLEAR" in "PARAMETER CLEAR" item of the parameter setting.
For further details on the parameter setting, refer to "5.2 Various Parameter Settings and
Setting Items" on page 63.
• When pressing
CANCEL
, only the previously input data is deleted. When it is pressed after comple-
tion of the subgroup measurement in WORK MODE "MODE3", the previously input subgroup will
be deleted.
For further details on the operation in WORK MODE "MODE3", refer to "3.3 Printing the
Data for Xbar-R Control Chart (MODE3)" on page 31.
• Be sure to delete the measurement data by pressing
CLEAR
before setting the limit data.
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED