EasyManuals Logo

Mitutoyo DP-1VA LOGGER User Manual

Mitutoyo DP-1VA LOGGER
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
44
3 Advanced Operations and Useful Functions
No. 99MAM029A
3
Start the measurement data input selecting the interval time.
Press the key to set the interval time.
» The interval time set and the starting
time/date are printed.
» Timer input starts.
Tips
The buzzer sounds at the time of measure-
ment data input but it does not when the
interval time set at 0.25 s.
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED
Set at 5 s
Set at 30 s
Set at 1 min
Set at 30 min
Set at 60 min
Set at 1 s
Interval time
Starting time
and date
Measure
-
ment data
Set at 0.25 s
4
Completion of the timer input.
Press
CLEAR
while pressing
PRINTER
.
» "Û INT TIMER OFF Û" and "Û PRINT-
ER ON Û" are printed.
Tips
In case 0.25 s or 1 s is set to the interval time, do not press
CANCEL
,
CLEAR
or
STAT.
OUT LOG
. It may
cause a malfunction.
If the data is already fetched when the timer input is completed, that data may be printed.
The interval time shall not be changed while the timer input of measurement data is performed.
The timer input shall be once terminated to change the interval time.
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mitutoyo DP-1VA LOGGER and is the answer not in the manual?

Mitutoyo DP-1VA LOGGER Specifications

General IconGeneral
BrandMitutoyo
ModelDP-1VA LOGGER
CategoryPrinter
LanguageEnglish