Command #175: Write Alarm Setpoint Control Actions
Writes the Ultima X Alarm Setpoint Control Actions. The Ultima X Gas
Monitor uses alarm setpoint Control Actions to enabled or disabled, set
to increasing or decreasing and to set the alarm to latching or non
latching.
Request Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Alarm Number (1, 2, or 3)
1 Bit Enum Alarm Control Action Value (see TABLE D-9)
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Alarm Number (1, 2, or 3)
1 Bit Enum Alarm Control Action Value (see TABLE D-9)
Command-Specific Response Codes
CODE CLASS DESCRIPTION
0 Success No Command-Specific Errors
1-4 Undefined
5 Error Too few data bytes
6 Undefined
7 Error In write protect mode
8-15 Undefined
16 Error Access Restricted
17-18 Undefined
19 Error Invalid device variable index
20-31 Undefined
32 Error Busy
33-127 Undefined