Command #181: Write Cal Signal Enable
This command writes command number to the Ultima X Gas Monitor to
enable or disable the Cal signal output. Without the Cal Signal enabled,
the 4-20 mA output will follow the gas reading during calibration. With
the Cal Signal enabled, the 4-20 mA output will be set to 3.75 ma during
calibration and be held there for one minute after calibration has ended
to allow the sensor to re-stabilize. Status group 3 indicates the current
setting of this mode.
Request Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Cal Signal 1 = Enable 0 = Disable
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Cal Signal 1 = Enable 0 = Disable
Command-Specific Response Codes
CODE CLASS DESCRIPTION
0 Success No Command-Specific Errors
1-2 Undefined
3 Error Parameter too large
4 Undefined
5 Error Too few data bytes
6 Undefined
7 Error In write protect mode
8-15 Undefined
16 Error Access Restricted
17-31 Undefined
32 Error Busy
33-127 Undefined
D-22