Command #134: Read Last Cal Date
Returns the Ultima last calibration date of the currently connected
sensor.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-2 Unsigned Last sensor calibration date
Command #135: Read Gas Table
This command returns the Ultima sensor Gas Table currently in use.
The Gas Tables are linearization reference tables used with certain
sensors to provide accurate response for different gases from the
same sensor.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Unsigned Gas Table Number (see TABLE D-10)
Command #136: Read Input Voltage Value
Returns the Ultima input supply voltage value. This number should be in
the range of 8-30 volts DC.
Request Data Bytes
None.
Response Data Bytes
BYTE FORMAT DESCRIPTION
0-3 Float Input Voltage Value
D-10