Route this cable with DC power to avoid noise interference
from relay wiring.
b. Connect the two-conductor cable to the two positions of
terminal block J10.
c. Identify the cable to enable proper connection at the
button.
d. Route cable to the push-button location; wire the button.
6. Re-install the wiring plugs.
Ensure that wiring plugs are firmly seated into their mates.
7. Pull the cable away from the unit to relieve any excess slack.
It is important not to have excess wire or cable within the
module to avoid unwanted AC noise.
8. Re-install the cover of the Ultima X Series Gas Monitor.
Figure A-2. Relay Printed Circuit Board
A-6