Command #180: Write Sensor Swap Delay Enable
This command writes command number to the Ultima X Gas Monitor to
enable or disable the two-minute swap delay feature. This device
feature enables a two-minute hold-off of the sensor missing fault,
allowing the user to “Swap” or change sensor modules without having
the 4-20 mA set to the fault condition. The configuration change bit will
be set, and the configuration change counter will be incremented.
Request Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Swap Delay 1 = Enable 0 = Disable
Response Data Bytes
BYTE FORMAT DESCRIPTION
0 Enum Swap Delay 1 = Enable 0 = Disable
Command-Specific Response Codes
CODE CLASS DESCRIPTION
0 Success No Command-Specific Errors
1-2 Undefined
3 Error Parameter too large
4 Undefined
5 Error Too few data bytes
6 Undefined
7 Error In write protect mode
8-15 Undefined
16 Error Access Restricted
17-31 Undefined
32 Error Busy
33-127 Undefined