EasyManua.ls Logo

Newport MM4005 - Page 141

Newport MM4005
417 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EDH0162En1040 06/99 3.58
MM4005 FT Set output frequency
Usage IMM PGM MIP
Syntax FTnn
Parameters
Description nn
[float] Desired frequency.
Range nn 0 and 0.01 to 5000.
Units nn
Hz.
Defaults nn Missing: 0.
Out of range: Error D.
Non-increment: Rounded to nearest frequency increment (see table).
Description This command controls the output signal on pin 24 of the 25-pin auxiliary
connector. The nn parameter represents the frequency of the output sig-
nal. Setting nn to 0 disables the frequency generator. The output has an
open collector configuration and a frequency range and resolution shown
in the following table:
F (Hz) F (Hz)
(Output frequency) (Frequency resolution)
0.01 - 20 0.001
20 - 250 0.010
250 - 500 0.020
500 - 1000 0.100
1000 - 2500 0.500
2500 - 5000 1.000
NOTE
For the hardware definition of the frequency generator port, please see
Appendix B, Connector Pinouts, Auxiliary Connector.
Returns
None.
Errors D Unauthorized execution.
Rel. Commands None.
Example
FT218.24
| Set an output frequency of 218.240 Hz on pin 24 of the auxiliary con-
nector.
.