7. Set the correct VS200 LED lamp voltages for the individual objectives. Refer
to chapter VS200 LED lamp voltages auf Seite 97 for detailed information.
18.6.2 Setup phase contrast (PH) observation method
If you want to add a phase contrast observation method do the following:
1. Select the [Acquire] >[Devices] > [Device Customization] command to
open the [Device Customization] dialog box.
2. Select the [BFMono] observation method.
3.
Use the [Copy Observation Method] button to make a copy of the
observation method.
4. Type in a name, e.g. "Phase Contrast".
5. In the [Type] picklist select [Phase Contrast].
6. Select the [BX3 Condenser] entry from the available components and
choose the [Adjust per objective] entry in the [Status] pick list.
- 92 -
Instructions for use
18 VS200 device configuration
OLYMPUS VS200 System