E-Drill General Fault
Trouble
Probable Cause
Remedy
E-Drill stops within a second of
starting, and does not go to
cut depth.
Ground Pin protrusion
significantly out of adjustment
or worn beyond its life.
Paint, corrosion, or sealant on
fastener head.
Adjust Ground Pin protrusion
with Ground Pin Setting Tool.
Clean fastener head for proper
grounding.
Cutting a depressed center
fastener (such as a Torx or
Phillips head) without specially
adjusting the Ground Pin
protrusion to match the
fastener depth.
Electrode advances to limit
and Hand Tool emits a
constant buzzing noise.
E-Drill Trigger is stuck in the
on position.
Power off system. Check
operation of trigger, clean
surrounding area to attempt to
free the trigger assembly.
Trigger should “clickwhen
depressed. Contact PPedm
Customer Service Hotline if
the problem persists.
Cutting is intermittent and
Electrode has come loose in
Remove Adapter, drive
cutting time is longer than
E-Drill.
electrode forward using the
usual.
trigger and re-tighten with the
electrode torque tool.
Water leaking from E-Drill
retract button or ground pin
adjusting holes.
E-Drill seal failure
Call the PPedm Customer
Service Hotline and obtain an
RMA and return E-Drill to
PPedm for service.
Vacuum pump does not stop
Water inside of E-Drill.
Call the PPedm Customer
operating after 3 seconds of
Service Hotline and obtain an
trigger release. E-Drill
RMA and return E-Drill to
“buzzing” and does not stop
PPedm for service.
automatically at the end of
stroke.