Section 8 Troubleshooting
I cannot connect to PT-Node
to change the stored network
settings
You can re-set the network settings of PT-Node to its
factory defaults. To do this disconnect the power supply
from PT-Node and wait a few seconds. Then re-connect
the power supply and wait until you see the status
screen again. Then again disconnect the power supply,
wait a few seconds, and re-connect it. Then do this a
third time. After the third re-start the display will show
“factory reset” and the network settings have been re-
set. Refer to the instructions above on how to perform
the initial network settings.
The PT-Node website does
not respond
Try to reload the page in your browser.
If this does not work, reboot PT-Node by unplugging the
power cable, waiting a few seconds, and plugging the
cable back in.
To prevent this from occurring do not disconnect or
connect any communication cable to PT-Node while the
instrument is powered on.