Technical Alarms (INOPs)
Troubleshooting 8-5
Alarm Text
Priority
Condition
What to do
Invalid Leadset
Source - MX40
Hard
FAST SpO2 leadset
attached to MX40
with Masimo SET
option.
Masimo SET leadset
attached to MX40
with FAST SpO2.
(MX40 Revision
dependent.)
Replace with compatible
leadset.
Leadset Life
Source - MX40
Soft
The single-patient use
leadset has exceeded
its limit of 25 cycles.
Replace with new leadset.
Leadset Unplugged
Source - MX40
Note — If supported by
your revision of PIC iX,
this INOP may also be
configured to display as a
Red or Yellow Technical
Alarm.
Note The message will
initially display for up to 3
seconds as “ECG Leads
Off
at the MX40 and at
the PIC iX.
Red or
Yellow or
Hard
Technical
Alarm
Patient cable has been
unplugged from the
MX40.
Incompatible leadset
attached to patient
cable. Masimo leadset
attached to MX40 with
FAST SpO2. (MX40
Revision dependent.)
Re-attach the patient cable.
Replace the leadset.
Replace with compatible
leadset.
Local Audio Off
Source - MX40
Note — This is normal
operation in Telemetry
Mode.
Soft There is no alarm audio
notification when
operating in Telemetry
Mode.
Change to Monitor Mode.
NBP Cuff Overpress
Source - Cableless
Measurement Device
Severe
Cuff pressure has
increased above
overpressure safety limits.
Remove cuff and tubing and
expel air.
NBP Equip Malf
Source - Cableless
Measurement Device
Hard Tubing may be
obstructed or kinked.
Hardware
malfunction.
Check tubing.
If condition persists,
contact Service.