Technical Alarms (INOPs)
Troubleshooting 8-5
•
Invalid Leadset
Source - MX40
Hard
• FAST SpO2 leadset
attached to MX40
with Masimo SET
option.
• Masimo SET leadset
attached to MX40
with FAST SpO2.
(MX40 Revision
dependent.)
Replace with compatible
leadset.
Leadset Life
Source - MX40
Soft
The single-patient use
leadset has exceeded
its limit of 25 cycles.
Replace with new leadset.
Leadset Unplugged
Source - MX40
Note — If supported by
your revision of PIC iX,
this INOP may also be
configured to display as a
Red or Yellow Technical
Alarm.
Note — The message will
initially display for up to 3
seconds as “ECG Leads
Off
” at the MX40 and at
the PIC iX.
Red or
Yellow or
Hard
Technical
Alarm
• Patient cable has been
unplugged from the
MX40.
• Incompatible leadset
attached to patient
cable. Masimo leadset
attached to MX40 with
FAST SpO2. (MX40
Revision dependent.)
• Re-attach the patient cable.
• Replace the leadset.
• Replace with compatible
leadset.
Local Audio Off
Source - MX40
Note — This is normal
operation in Telemetry
Mode.
Soft There is no alarm audio
notification when
operating in Telemetry
Mode.
Change to Monitor Mode.
NBP Cuff Overpress
Source - Cableless
Measurement Device
Severe
Cuff pressure has
increased above
overpressure safety limits.
Remove cuff and tubing and
expel air.
NBP Equip Malf
Source - Cableless
Measurement Device
Hard • Tubing may be
obstructed or kinked.
• Hardware
malfunction.
• Check tubing.
• If condition persists,
contact Service.