Technical Alarms (INOPs)
8-4 Troubleshooting
Alarm Text
Priorit
Condition
What to do
cl SpO2 Batt Empty
Source - Cableless
Measurement Device
Hard CL SpO
2
Pod
empty battery
condition.
Monitoring is not
possible.
Replace CL SpO
2
Pod.
Recharge depleted SpO
2
Pod.
cl SpO2 Batt Low
Source - Cableless
Measurement Device
Hard
CL SpO
2
Pod
weak battery
condition.
Charge CL SpO
2
Pod.
cl SpO2 Disconnect
Source - Cableless
Measurement Device
Hard CL SpO
2
Pod is
not connected
with the MX40.
Resolve
interference
condition.
Reduce range between
CL
SpO
2
Pod and
MX40.
Cuff Not Deflated
Source - Cableless
Measurement Device
Severe Cuff pressure has
exceeded the
specified safety limit.
Remove cuff and tubing
and expel air.
ECG/Arrh AlarmsOff
Source - MX40
Soft
ECG is turned off.
This is normal
behavior if operating
in SpO
2
only
mode.
Turn on ECG.
Attach ECG/SpO
2
patient
cable.
<ECG Lead>Lead Off
Source - MX40
Hard Single lead is
off.
If primary lead is MCL,
lead will be identified
as V/C in INOP text.
Re-attach ECG leads
to patient.
ECG Leads Off
Note — This INOP may
also be configured to
display as a Red or
Yellow Technical Alarm.
Source - MX40
Red or
Y
e
llo
w
or Hard
Technic
al
Alarm
Multiple leads are off. Re-attach ECG leads
to patient.